Home —» News —» Corrosion gauge vs flaw detector: when you need both

Corrosion gauge vs flaw detector: when you need both

Dakota NDT family of ultrasonic instruments for thickness and corrosion testing

A corrosion gauge and a flaw detector use the same ultrasonic physics but solve different problems: measuring thickness versus locating internal discontinuities. A guide to understanding when one of the two is enough, when you need both and when to choose a hybrid Dakota FX instrument.

Many NDT inspectors wonder whether to buy a corrosion gauge or a flaw detector, or both. The answer depends on the application: the two instruments use the same ultrasonic physics but are optimised for different tasks. A corrosion gauge is designed to measure thickness accurately and document degradation over time; a flaw detector is designed to locate, size and classify internal discontinuities in materials.

Dakota corrosion gauges

What a corrosion gauge does

A corrosion gauge (thickness gauge) is an instrument dedicated to measuring the remaining thickness of metal walls, typically to monitor the progress of corrosion. The Dakota NDT models of the CX, ZX, MX, CMX series are optimised for this purpose with specific features.

  • Dual element probes: two piezoelectric crystals (separate transmitter and receiver) that allow reliable measurements even on corroded and irregular surfaces
  • Echo-Echo ThruPaint™ mode: ignores coatings up to 2 mm and measures only the thickness of the metal
  • High-speed scan mode (up to 250 readings/sec on the MX and CMX models) for mapping large surfaces quickly
  • Data logger structured for repetitive inspections with grids and batches
  • Hi/Lo alarms to flag automatically any thickness below the critical threshold
  • Differential mode to compare the current measurement with a reference value
  • Compliance with inspection standards: ASTM E 797, EN 14127, EN 15317

What a flaw detector does

Dakota FX flaw detector

A flaw detector is an instrument that displays the ultrasonic waveform as an A-Scan and lets the operator identify, measure and classify internal discontinuities: cracks, inclusions, porosity, delaminations. The Dakota FX70-DL, FX71-DL and FX81-DL models are designed for this purpose with specific functions.

  • Sizing toolkit: DAC, AWS, TCG, DGS for sizing defects according to recognised standards
  • Angle shear wave probes (in addition to single element probes) for weld inspection
  • PRF adjustable up to 2000 Hz for fast scanning
  • 3 detection modes: Z-Cross, Flank, Peak — to choose the measurement point on the echo
  • TCG (Time Corrected Gain): automatically compensates for attenuation with depth
  • A-Scan and B-Scan display for graphical interpretation of the signal
  • EN 12668-1 compliance for flaw detectors
  • Handling of wedges and angle transducers for the inspection of welded joints

The key practical differences

AspectCorrosion gaugeFlaw detector
Primary purposeMeasure remaining thicknessLocate and size defects
Typical probesDual element, corrosion probesSingle element and angle shear wave
DisplayNumeric + B-ScanDetailed A-Scan + B-Scan
Sizing toolkit—DAC, AWS, TCG, DGS
Typical scan mode250 readings/secPRF up to 2000 Hz
Data loggerYes, structured in gridsYes, with A/B-Scan images
Reference standardASTM E 797, EN 14127EN 12668-1
Weld inspectionNot suitableTypical application
Corrosion monitoringTypical applicationPossible but oversized

When you need both

Inspection of pipes and tanks with NDT instruments

There are situations in which a company or an inspector needs both instruments in their kit:

  • Maintenance of process plants: the corrosion gauge is used for periodic thickness checks on piping and tanks (API 570, API 653); the flaw detector is used to inspect welds during repairs or modifications
  • New construction and acceptance testing: the flaw detector is mandatory for weld quality control (AWS D1.1, EN ISO 17640); the corrosion gauge is used for baseline dimensional checks
  • NDT inspection contractors: anyone offering services must be able to respond to varied requests from customers
  • Aerospace and composites: detecting delaminations calls for a flaw detector, while thickness checks on machined components call for a precision gauge

Dakota NDT hybrid alternatives

The FX70-DL, FX71-DL and FX81-DL models of the FX series combine both functions in a single instrument: they are fully fledged flaw detectors, but can be switched to thickness gauge mode to measure material and coating thickness simultaneously (Pulse-Echo Coating Mode). They are a sound solution when working in mixed contexts, but they do not replace the specialisation of a dedicated gauge for intensive thickness mapping campaigns.


To assess with us the configuration best suited to Your needs — dedicated gauge, dedicated detector or hybrid solution — contact our technical office.