The Dakota PMX2-DL is a precision thickness gauge: selectable resolution down to 0.001 mm, five measurement modes with delay line and pencil probes, A-Scan and B-Scan. It is the instrument for the world of machining, not of corrosion.
- ✓ Selectable resolution 0.01 or 0.001 mm
- ✓ Five modes: P-E, E-E, E-E Verify, Interface-Echo and Interface-Echo-Echo
- ✓ Single contact, delay line and pencil probes from 1 to 25 MHz
- ✓ 100, 150 and 200 V pulser with spike, thin and wide pulse width
- ✓ 110 dB gain, TDG, damping over six values (35–1500 ohm)
- ✓ Three gates, adjustable, 64 setups, scan at 250 readings/s
- ✓ A-Scan and B-Scan, 4 GB memory, USB-C
Measuring corrosion and measuring with precision are two different trades. In the first, what matters is the minimum remaining wall on a thick, irregular surface, and a hundredth of a millimetre does not change the decision. In the second — a pressed sheet, a bottle wall, an aerospace component in titanium — the micron is exactly what separates the conforming part from the reject.
On this page
The PMX2-DL belongs to the second world. It works with delay line and pencil probes up to 25 MHz and introduces two modes that corrosion gauges do not have: Interface-Echo and Interface-Echo-Echo, which take as their reference the interface between the delay line and the part. It is the way to go below the millimetre without confusing the entry echo with the backwall echo.

Need help choosing the right product?
Compared with the PMX3-DL only the display changes: greyscale here, colour AMOLED there. Everything else — electronics, modes, range, gates, memory — is identical, and in greyscale the battery life is almost three times as long.
Position within the precision range
| Model | Specialisation | Display |
|---|---|---|
| PZX1-DL | thin substrates | LCD |
| PMX2-DL | universal precision with A-Scan | greyscale VGA |
| PMX3-DL | universal precision | colour AMOLED |
| PMX4-DL | high penetration, attenuating materials | see datasheet |
| PCX8-DL | precision thickness gauge | see datasheet |
For the complete picture of the family: Dakota PMX2-DL / PMX3-DL.
When to choose the PMX2-DL

- Thin walls, where the micron matters and not the tenth
- Pressed sheet, deep-drawn parts, container walls
- Aluminium and titanium components for aerospace
- Plastics: Interface-Echo mode starts from 0.127 mm
- Quality control in production with tight tolerances
- Precision measurements have to be stored and reported
The five measurement modes
| Mode | Probe | Range on steel |
|---|---|---|
| P-E — Pulse-Echo | contact | 1 mm – 9.14 m |
| E-E — Echo-Echo | contact | 2.54 – 914.4 mm |
| E-EV — Echo-Echo Verify | contact | 2.54 – 152.4 mm |
| I-E — Interface-Echo | delay line and pencil | 1.27 – 25.4 mm; plastics from 0.127 mm |
| I-E-E — Interface-Echo-Echo | delay line and pencil | 0.152 – 12.7 mm |
The two Interface modes are what sets a precision gauge apart from a corrosion one: they take as their reference the echo at the interface between the delay line and the part, making it possible to measure thicknesses that in Pulse-Echo would be indistinguishable from the entry echo.
Main features
Fine control of the signal
- Pulser adjustable to 100, 150 and 200 V
- Variable pulse width: spike, thin, wide
- Gain over 110 dB, manual or AGC according to the mode
- TDG with variable start and slope
- Damping over six values: 35, 50, 75, 300, 600 and 1500 ohm
- PRF up to 250 Hz, 100 MHz DSP platform on FPGA
Working functions
- Three adjustable gates, according to the selected mode
- 64 custom setups defined by the user
- Scan mode: 250 readings per second, shows the minimum when the probe is lifted
- Alarm with minimum and maximum thresholds, audible and LED indication
- OBSTRUCT function for inaccessible points
Display and storage
- Rectified and RF A-Scan, 25 Hz refresh
- Cross-sectional B-Scan, from 10 to 200 readings/s
- 4 GB internal memory, alphanumeric grid or sequential log
- Screen capture in bitmap (.tif)
- USB-C output, Windows interface software
Full technical specifications
Data taken from the manufacturer’s official datasheet.
| Parameter | Value |
|---|---|
| Part Number | Z-157-0000 |
| Measuring mode | P-E, E-E, E-EV, I-E, I-E-E |
| Resolution | selectable: ± 0.01 mm or ± 0.001 mm |
| Velocity range | 309.88 – 18,542 m/s |
| Calibration | 1 or 2 point, or selection from a material list |
| Platform | 100 MHz DSP on FPGA technology |
| Pulser | square wave, 100 / 150 / 200 V; spike, thin, wide width |
| Receiver | manual gain or AGC over a 110 dB range |
| Damping | 35, 50, 75, 300, 600, 1500 ohm |
| PRF | up to 250 Hz |
| Timing | precision TCXO, 100 MHz 8-bit digitizer |
| Gates | three, adjustable according to the mode |
| Setups | 64 custom |
| Transducers | single contact, delay line and pencil, 1 – 25 MHz |
| Connectors | quick-release LEMO 00; standard 4-foot cable |
| Display | greyscale VGA 1/8 in, 240 × 160 pixels, EL backlight with inversion |
| Views | rectified and RF A-Scan (25 Hz), B-Scan, large digits |
| Keypad | membrane, 12 keys with tactile feedback |
| Memory | 4 GB internal; alphanumeric grid or sequential log |
| Data output | USB-C, Windows interface software |
| Power supply | three AA cells, or USB-C from a PC or a socket |
| Battery life | alkaline 35 h, NiCad 10 h, NiMH 35 h |
| Operating temperature | −10 / +60 °C |
| Housing | extruded aluminium, nickel-plated aluminium end caps, gasket sealed |
| Weight | 383 g (13.5 oz) with batteries |
| Dimensions | 63.5 × 165 × 31.5 mm |
Typical applications
Aerospace and precision components
Measurement on aluminium and titanium with tight tolerances: see NDT in aerospace: precision measurement on aluminium and titanium components.
Pressed sheet and deep-drawn parts
Checking the thinning in the areas of greatest deformation, where conventional dimensional inspection cannot reach because the part is closed.
Plastics
In Interface-Echo mode measurement on plastics starts from 0.127 mm. For the method on non-metallic materials: How to improve results on difficult materials.
Quality control in production
With threshold alarms and storage of the readings: see Thickness measurement in incoming goods quality control.
Order codes: instrument, probes and accessories
Below you will find everything you can order for the PMX2-DL: the instrument, compatible probes, couplants, calibration blocks and accessories. Please quote the order code in your request: it is the code shown on the quotation and on the order confirmation.
The probe is not included with the instrument: it must be chosen and ordered separately from those listed on this page. All probes in the table are compatible with the PMX2-DL.
Instrument
Supplied with: PMX2-DL instrument, ultrasonic couplant, manual, calibration certificate, AA batteries, plastic carrying case. Probe not included.
| Image | Order code | SKU | Description |
|---|---|---|---|
![]() | Z-157-0000 | 10-15-62 | PMX2-DL precision thickness gauge with A-scan, data logger Probe to be ordered separately |
Compatible probes
Choose the frequency according to thickness and material: low frequencies for thick walls or attenuating materials, 5 MHz for general use, high frequencies for thin sections and small surfaces. A smaller diameter makes coupling on curved surfaces easier.
| Image | Order code | SKU | Description |
|---|---|---|---|
![]() | T-402-5507 | 14-26-04 | Single-element probe a linea di ritardo Standard, 15 MHz, Ø 1/4″, Microdot connector, side exit |
![]() | T-402-6507 | 10-15-67 | Single-element probe a linea di ritardo Standard, 20 MHz, Ø 1/4″, Microdot connector, side exit |
![]() | T-481-4507 | 14-20-93 | Single-element probe pencil, punta 1/16″, 10 MHz, Ø 3/16″, Microdot connector, top exit |
![]() | T-581-5507 | 10-15-68 | Single-element probe pencil, punta 1/16″, 15 MHz, Ø 3/16″, Microdot connector, top exit |
![]() | T-681-4507 | 10-14-00 | Single-element probe pencil a 90°, 10 MHz, Ø 3/16″, Microdot connector, side exit |
![]() | T-4033-1855 | 10-15-69 | Single-element contact probe Finger Tip, 2.25 MHz, Ø 3/8″, connector Microdot, side exit |
![]() | T-4023-1855 | 10-15-70 | Single-element contact probe Finger Tip, 2.25 MHz, Ø 1/4″, connector Microdot, side exit |
![]() | T-4043-1855 | 10-15-71 | Single-element contact probe Finger Tip, 2.25 MHz, Ø 1/2″, connector Microdot, side exit |
![]() | T-4023-2855 | 10-15-56 | Single-element contact probe Finger Tip, 5 MHz, Ø 1/4″, connector Microdot, side exit |
![]() | T-4033-2855 | 10-15-72 | Single-element contact probe Finger Tip, 5 MHz, Ø 3/8″, connector Microdot, side exit |
![]() | T-4043-2855 | 10-15-73 | Single-element contact probe Finger Tip, 5 MHz, Ø 1/2″, connector Microdot, side exit |
![]() | T-4023-4855 | 10-15-57 | Single-element contact probe Finger Tip, 10 MHz, Ø 1/4″, Microdot connector, side exit |
![]() | T-4033-4855 | 10-15-74 | Single-element contact probe Finger Tip, 10 MHz, Ø 3/8″, Microdot connector, side exit |
![]() | T-5903-2875 | 10-15-75 | Single-element contact probe Finger Tip Slim Line (non-magnetic), 5 MHz, Ø 1/8″, connector Microdot, top exit |
![]() | T-5903-4875 | 10-15-76 | Single-element contact probe Finger Tip Slim Line, 10 MHz, Ø 1/8″, connector Microdot, top exit |
Delay lines
Replacement delay lines for delay-line probes: choose the material and the diameter of the probe in use.
| Image | Order code | SKU | Description |
|---|---|---|---|
![]() | F-000-7102 | 10-15-59 | Standard delay line, 1/4″ x 3/8″ L, acrylic |
![]() | F-000-7103 | 10-15-60 | Standard delay line, 1/4″ x 1/2″ L, acrylic |
![]() | X-132-0000 | 14-17-32 | Linea di ritardo conica, 1/8″, acrylic |
![]() | X-532-0000 | 14-12-83 | Linea di ritardo conica, 3/16″, acrylic |
![]() | X-633-0000 | 14-09-20 | Standard delay line, 1/4″, graphite |
![]() | X-533-0000 | 14-10-63 | Linea di ritardo conica, 3/16″, graphite |
![]() | X-152-0000 | 14-24-17 | Delay line for pencil probe, 1/16″ x 0.45 L, acrylic |
![]() | X-752-0000 | 14-27-49 | Delay line for pencil probe, 1/8″ x 0.45 L, acrylic |
Probe cables
Cables connecting single-element probes to the instrument.
| Image | Order code | SKU | Description |
|---|---|---|---|
![]() | N-104-0020 | 14-16-00 | Probe cable 1.2 m single Lemo 00 – Microdot |
![]() | N-110-0020 | 14-04-08 | Probe cable 3 m single Lemo – Microdot |
![]() | N-120-0020 | 10-15-61 | Probe cable 6 m single Lemo – Microdot |
Couplants
A couplant is included in the supply; for refills and for high temperatures, choose from the following.
| Image | Order code | SKU | Description |
|---|---|---|---|
![]() | GEL-ACC-EXT | 12-13-51 | Blue anti-corrosion couplant gel, RCCM compliant. Extended temperature range -20 / 120 °C. 500 ml pack |
![]() | PST-ACC-MHT | 12-13-52 | Special medium-temperature ultrasonic coupling paste -30 / +300 °C. Non-flammable and smoke-free. RCCM compliant. 170 g pack |
![]() | PST-ACC-UHT | 12-13-53 | Special very-high-temperature ultrasonic coupling paste +100 / 520 °C. Non-flammable and smoke-free. RCCM compliant. 200 g pack |
Calibration blocks
For periodic verification of the instrument: choose the step block covering the thickness range you measure.
| Image | Order code | SKU | Description |
|---|---|---|---|
![]() | X-000-0001 | 10-15-48 | 5-step calibration block 0.1″–0.5″, steel 4340 |
![]() | X-000-0002 | 10-15-49 | 4-step calibration block 0.25″–1.0″, steel 4340 |
![]() | X-000-0022 | 10-15-50 | 5-step calibration block 2.5–12.5 mm, steel 1018 |
![]() | X-000-0021 | 10-15-51 | 4-step calibration block 6.25–25 mm, steel 1018 |
![]() | X-000-0005 | 10-15-52 | Case for calibration block |
Accessories and spare parts
Cases, cables and accessories specific to this instrument.
| Image | Order code | SKU | Description |
|---|---|---|---|
![]() | F-149-0001 | 10-15-64 | Nylon instrument case (MX, PMX, CMX, MVX, PVX, MMX, FX71, BT1 / MINI-MAX) Case only: instrument NOT included |
![]() | A-149-6002 | 10-15-65 | Rubber instrument boot (MX, PMX, CMX, MVX, PVX, MMX, FX71, BT1- DL v2.0) Case only: instrument NOT included |
![]() | A-100-6003 | 10-15-66 | Plastic carrying case (MX, PMX, CMX, MVX, PVX, MMX, FX71, BT1) Case only: instrument NOT included |
![]() | N-003-0330 | 10-15-39 | USB-C / USB-A data cable 0.9 m (V2.0 instruments) |
![]() | F-000-7000 | 10-15-77 | Spring-loaded bell probe holder (V-Block) – pencil probes, top cable exit models only |
For more on choosing the probe: probes for precision gauges and the guide How to choose the right ultrasonic transducer. For a quotation, please state the order code and quantity.
Documentation download
Official Dakota NDT technical documentation. The file opens in a new tab.
| Datasheet Dakota PMX2-DL and PMX3-DL Manufacturer’s official datasheet — PDF, in English |
Standards and certifications
- Factory calibration NIST-traceable and compliant with MIL-STD-45662A
- Suitable for measurements to EN 14127 and ASTM E797
For the regulatory picture: NDT standards for thickness measurement.
Frequently asked questions
The field of use and the type of probe. A corrosion gauge uses dual element probes and is optimised for thick, irregular walls, with resolution to the hundredth. A precision one such as the PMX2-DL uses single element, delay line and pencil probes, goes down to 0.001 mm of resolution and offers the Interface-Echo modes, which are essential on thin walls.
In Interface-Echo-Echo mode the range on steel starts from 0.152 mm; in Interface-Echo, measurement on plastics starts from 0.127 mm. The figures depend on the probe used and on the material.
Only the display: greyscale on the PMX2-DL, colour AMOLED at 60 Hz on the PMX3-DL. Electronics, modes, range, gates and memory are identical. In greyscale the battery life is 35 hours against 12.
In contact Pulse-Echo mode the range reaches 9.14 metres on steel, so technically yes. For corrosion campaigns, however, the instruments of the CMX series are more suitable: they use dual element probes and offer the through-coating modes.
For a quotation or a trial on your own parts, contact our technical office.
Further information and contacts
Systems division and sensor integration partner : www.roder.it
More information about RODER SRL : about us
Contact for general information : info@roder.it
Contact for information about these products : info@roder.it
The information contained on this web page is provided for information purposes only. Although it has been compiled with the utmost care, it does not constitute a contractual offer or a binding commitment to supply. It may contain transcription, translation or typographical errors. For precise and up-to-date information, please contact our company directly.
Some images on this website have been intentionally generated using Artificial Intelligence (AI). This is because, for many applications and projects, it is not possible to disclose photographs of the installation or of the actual system due to confidentiality agreements, confidentiality clauses and non-disclosure agreements (NDAs).