Home —» Products —» Thickness Gauges —» Dakota PMX2-DL – Precision gauge with A-Scan

Dakota PMX2-DL – Precision gauge with A-Scan

Dakota NDT PMX2-DL precision thickness gauge with A-Scan display

The Dakota PMX2-DL is a precision thickness gauge: selectable resolution down to 0.001 mm, five measurement modes with delay line and pencil probes, A-Scan and B-Scan. It is the instrument for the world of machining, not of corrosion.

  • ✓ Selectable resolution 0.01 or 0.001 mm
  • ✓ Five modes: P-E, E-E, E-E Verify, Interface-Echo and Interface-Echo-Echo
  • ✓ Single contact, delay line and pencil probes from 1 to 25 MHz
  • ✓ 100, 150 and 200 V pulser with spike, thin and wide pulse width
  • ✓ 110 dB gain, TDG, damping over six values (35–1500 ohm)
  • ✓ Three gates, adjustable, 64 setups, scan at 250 readings/s
  • ✓ A-Scan and B-Scan, 4 GB memory, USB-C

Measuring corrosion and measuring with precision are two different trades. In the first, what matters is the minimum remaining wall on a thick, irregular surface, and a hundredth of a millimetre does not change the decision. In the second — a pressed sheet, a bottle wall, an aerospace component in titanium — the micron is exactly what separates the conforming part from the reject.

The PMX2-DL belongs to the second world. It works with delay line and pencil probes up to 25 MHz and introduces two modes that corrosion gauges do not have: Interface-Echo and Interface-Echo-Echo, which take as their reference the interface between the delay line and the part. It is the way to go below the millimetre without confusing the entry echo with the backwall echo.

Compared with the PMX3-DL only the display changes: greyscale here, colour AMOLED there. Everything else — electronics, modes, range, gates, memory — is identical, and in greyscale the battery life is almost three times as long.

Position within the precision range

ModelSpecialisationDisplay
PZX1-DLthin substratesLCD
PMX2-DLuniversal precision with A-Scangreyscale VGA
PMX3-DLuniversal precisioncolour AMOLED
PMX4-DLhigh penetration, attenuating materialssee datasheet
PCX8-DLprecision thickness gaugesee datasheet

For the complete picture of the family: Dakota PMX2-DL / PMX3-DL.

When to choose the PMX2-DL

PMX2-DL precision thickness gauge with A-scan, data logger
  • Thin walls, where the micron matters and not the tenth
  • Pressed sheet, deep-drawn parts, container walls
  • Aluminium and titanium components for aerospace
  • Plastics: Interface-Echo mode starts from 0.127 mm
  • Quality control in production with tight tolerances
  • Precision measurements have to be stored and reported

The five measurement modes

ModeProbeRange on steel
P-E — Pulse-Echocontact1 mm – 9.14 m
E-E — Echo-Echocontact2.54 – 914.4 mm
E-EV — Echo-Echo Verifycontact2.54 – 152.4 mm
I-E — Interface-Echodelay line and pencil1.27 – 25.4 mm; plastics from 0.127 mm
I-E-E — Interface-Echo-Echodelay line and pencil0.152 – 12.7 mm

The two Interface modes are what sets a precision gauge apart from a corrosion one: they take as their reference the echo at the interface between the delay line and the part, making it possible to measure thicknesses that in Pulse-Echo would be indistinguishable from the entry echo.

Main features

Fine control of the signal

  • Pulser adjustable to 100, 150 and 200 V
  • Variable pulse width: spike, thin, wide
  • Gain over 110 dB, manual or AGC according to the mode
  • TDG with variable start and slope
  • Damping over six values: 35, 50, 75, 300, 600 and 1500 ohm
  • PRF up to 250 Hz, 100 MHz DSP platform on FPGA

Working functions

  • Three adjustable gates, according to the selected mode
  • 64 custom setups defined by the user
  • Scan mode: 250 readings per second, shows the minimum when the probe is lifted
  • Alarm with minimum and maximum thresholds, audible and LED indication
  • OBSTRUCT function for inaccessible points

Display and storage

  • Rectified and RF A-Scan, 25 Hz refresh
  • Cross-sectional B-Scan, from 10 to 200 readings/s
  • 4 GB internal memory, alphanumeric grid or sequential log
  • Screen capture in bitmap (.tif)
  • USB-C output, Windows interface software

Full technical specifications

Data taken from the manufacturer’s official datasheet.

ParameterValue
Part NumberZ-157-0000
Measuring modeP-E, E-E, E-EV, I-E, I-E-E
Resolutionselectable: ± 0.01 mm or ± 0.001 mm
Velocity range309.88 – 18,542 m/s
Calibration1 or 2 point, or selection from a material list
Platform100 MHz DSP on FPGA technology
Pulsersquare wave, 100 / 150 / 200 V; spike, thin, wide width
Receivermanual gain or AGC over a 110 dB range
Damping35, 50, 75, 300, 600, 1500 ohm
PRFup to 250 Hz
Timingprecision TCXO, 100 MHz 8-bit digitizer
Gatesthree, adjustable according to the mode
Setups64 custom
Transducerssingle contact, delay line and pencil, 1 – 25 MHz
Connectorsquick-release LEMO 00; standard 4-foot cable
Displaygreyscale VGA 1/8 in, 240 × 160 pixels, EL backlight with inversion
Viewsrectified and RF A-Scan (25 Hz), B-Scan, large digits
Keypadmembrane, 12 keys with tactile feedback
Memory4 GB internal; alphanumeric grid or sequential log
Data outputUSB-C, Windows interface software
Power supplythree AA cells, or USB-C from a PC or a socket
Battery lifealkaline 35 h, NiCad 10 h, NiMH 35 h
Operating temperature−10 / +60 °C
Housingextruded aluminium, nickel-plated aluminium end caps, gasket sealed
Weight383 g (13.5 oz) with batteries
Dimensions63.5 × 165 × 31.5 mm

↑ Back to index

Typical applications

Aerospace and precision components

Measurement on aluminium and titanium with tight tolerances: see NDT in aerospace: precision measurement on aluminium and titanium components.

Pressed sheet and deep-drawn parts

Checking the thinning in the areas of greatest deformation, where conventional dimensional inspection cannot reach because the part is closed.

Plastics

In Interface-Echo mode measurement on plastics starts from 0.127 mm. For the method on non-metallic materials: How to improve results on difficult materials.

Quality control in production

With threshold alarms and storage of the readings: see Thickness measurement in incoming goods quality control.

Order codes: instrument, probes and accessories

Below you will find everything you can order for the PMX2-DL: the instrument, compatible probes, couplants, calibration blocks and accessories. Please quote the order code in your request: it is the code shown on the quotation and on the order confirmation.

The probe is not included with the instrument: it must be chosen and ordered separately from those listed on this page. All probes in the table are compatible with the PMX2-DL.

Instrument

Supplied with: PMX2-DL instrument, ultrasonic couplant, manual, calibration certificate, AA batteries, plastic carrying case. Probe not included.

ImageOrder codeSKUDescription
PMX2-DL precision thickness gauge with A-scan, data loggerZ-157-000010-15-62PMX2-DL precision thickness gauge with A-scan, data logger
Probe to be ordered separately

Compatible probes

Choose the frequency according to thickness and material: low frequencies for thick walls or attenuating materials, 5 MHz for general use, high frequencies for thin sections and small surfaces. A smaller diameter makes coupling on curved surfaces easier.

ImageOrder codeSKUDescription
Single-element probe a linea di ritardo Standard, 15 MHz, Ø 1/4T-402-550714-26-04Single-element probe a linea di ritardo Standard, 15 MHz, Ø 1/4″, Microdot connector, side exit
Single-element probe a linea di ritardo Standard, 20 MHz, Ø 1/4T-402-650710-15-67Single-element probe a linea di ritardo Standard, 20 MHz, Ø 1/4″, Microdot connector, side exit
Single-element probe pencil, punta 1/16T-481-450714-20-93Single-element probe pencil, punta 1/16″, 10 MHz, Ø 3/16″, Microdot connector, top exit
Single-element probe pencil, punta 1/16T-581-550710-15-68Single-element probe pencil, punta 1/16″, 15 MHz, Ø 3/16″, Microdot connector, top exit
Single-element probe pencil a 90°, 10 MHz, Ø 3/16T-681-450710-14-00Single-element probe pencil a 90°, 10 MHz, Ø 3/16″, Microdot connector, side exit
Single-element contact probe Finger Tip, 2.25 MHz, Ø 3/8T-4033-185510-15-69Single-element contact probe Finger Tip, 2.25 MHz, Ø 3/8″, connector Microdot, side exit
Single-element contact probe Finger Tip, 2.25 MHz, Ø 1/4T-4023-185510-15-70Single-element contact probe Finger Tip, 2.25 MHz, Ø 1/4″, connector Microdot, side exit
Single-element contact probe Finger Tip, 2.25 MHz, Ø 1/2T-4043-185510-15-71Single-element contact probe Finger Tip, 2.25 MHz, Ø 1/2″, connector Microdot, side exit
Single-element contact probe Finger Tip, 5 MHz, Ø 1/4T-4023-285510-15-56Single-element contact probe Finger Tip, 5 MHz, Ø 1/4″, connector Microdot, side exit
Single-element contact probe Finger Tip, 5 MHz, Ø 3/8T-4033-285510-15-72Single-element contact probe Finger Tip, 5 MHz, Ø 3/8″, connector Microdot, side exit
Single-element contact probe Finger Tip, 5 MHz, Ø 1/2T-4043-285510-15-73Single-element contact probe Finger Tip, 5 MHz, Ø 1/2″, connector Microdot, side exit
Single-element contact probe Finger Tip, 10 MHz, Ø 1/4T-4023-485510-15-57Single-element contact probe Finger Tip, 10 MHz, Ø 1/4″, Microdot connector, side exit
Single-element contact probe Finger Tip, 10 MHz, Ø 3/8T-4033-485510-15-74Single-element contact probe Finger Tip, 10 MHz, Ø 3/8″, Microdot connector, side exit
Single-element contact probe Finger Tip Slim Line (non-magnetic), 5 MHz, Ø 1/8T-5903-287510-15-75Single-element contact probe Finger Tip Slim Line (non-magnetic), 5 MHz, Ø 1/8″, connector Microdot, top exit
Single-element contact probe Finger Tip Slim Line, 10 MHz, Ø 1/8T-5903-487510-15-76Single-element contact probe Finger Tip Slim Line, 10 MHz, Ø 1/8″, connector Microdot, top exit

Delay lines

Replacement delay lines for delay-line probes: choose the material and the diameter of the probe in use.

ImageOrder codeSKUDescription
Standard delay line, 1/4F-000-710210-15-59Standard delay line, 1/4″ x 3/8″ L, acrylic
Standard delay line, 1/4F-000-710310-15-60Standard delay line, 1/4″ x 1/2″ L, acrylic
Linea di ritardo conica, 1/8X-132-000014-17-32Linea di ritardo conica, 1/8″, acrylic
Linea di ritardo conica, 3/16X-532-000014-12-83Linea di ritardo conica, 3/16″, acrylic
Standard delay line, 1/4X-633-000014-09-20Standard delay line, 1/4″, graphite
Linea di ritardo conica, 3/16X-533-000014-10-63Linea di ritardo conica, 3/16″, graphite
Delay line for pencil probe, 1/16X-152-000014-24-17Delay line for pencil probe, 1/16″ x 0.45 L, acrylic
Delay line for pencil probe, 1/8X-752-000014-27-49Delay line for pencil probe, 1/8″ x 0.45 L, acrylic

Probe cables

Cables connecting single-element probes to the instrument.

ImageOrder codeSKUDescription
Probe cable 1.2 m single Lemo 00 – MicrodotN-104-002014-16-00Probe cable 1.2 m single Lemo 00 – Microdot
Probe cable 3 m single Lemo – MicrodotN-110-002014-04-08Probe cable 3 m single Lemo – Microdot
Probe cable 6 m single Lemo – MicrodotN-120-002010-15-61Probe cable 6 m single Lemo – Microdot

Couplants

A couplant is included in the supply; for refills and for high temperatures, choose from the following.

ImageOrder codeSKUDescription
Blue anti-corrosion couplant gel, RCCM compliant. Extended temperature range -20 / 120 °C. 500 ml packGEL-ACC-EXT12-13-51Blue anti-corrosion couplant gel, RCCM compliant. Extended temperature range -20 / 120 °C. 500 ml pack
Special medium-temperature ultrasonic coupling paste -30 / +300 °C. Non-flammable and smoke-free. RCCM compliant. 170 g packPST-ACC-MHT12-13-52Special medium-temperature ultrasonic coupling paste -30 / +300 °C. Non-flammable and smoke-free. RCCM compliant. 170 g pack
Special very-high-temperature ultrasonic coupling paste +100 / 520 °C. Non-flammable and smoke-free. RCCM compliant. 200 g packPST-ACC-UHT12-13-53Special very-high-temperature ultrasonic coupling paste +100 / 520 °C. Non-flammable and smoke-free. RCCM compliant. 200 g pack

Calibration blocks

For periodic verification of the instrument: choose the step block covering the thickness range you measure.

ImageOrder codeSKUDescription
5-step calibration block 0.1X-000-000110-15-485-step calibration block 0.1″–0.5″, steel 4340
4-step calibration block 0.25X-000-000210-15-494-step calibration block 0.25″–1.0″, steel 4340
5-step calibration block 2.5–12.5 mm, steel 1018X-000-002210-15-505-step calibration block 2.5–12.5 mm, steel 1018
4-step calibration block 6.25–25 mm, steel 1018X-000-002110-15-514-step calibration block 6.25–25 mm, steel 1018
Case for calibration blockX-000-000510-15-52Case for calibration block

Accessories and spare parts

Cases, cables and accessories specific to this instrument.

ImageOrder codeSKUDescription
Nylon instrument case (MX, PMX, CMX, MVX, PVX, MMX, FX71, BT1 / MINI-MAX)F-149-000110-15-64Nylon instrument case (MX, PMX, CMX, MVX, PVX, MMX, FX71, BT1 / MINI-MAX)
Case only: instrument NOT included
Rubber instrument boot (MX, PMX, CMX, MVX, PVX, MMX, FX71, BT1- DL v2.0)A-149-600210-15-65Rubber instrument boot (MX, PMX, CMX, MVX, PVX, MMX, FX71, BT1- DL v2.0)
Case only: instrument NOT included
Plastic carrying case (MX, PMX, CMX, MVX, PVX, MMX, FX71, BT1)A-100-600310-15-66Plastic carrying case (MX, PMX, CMX, MVX, PVX, MMX, FX71, BT1)
Case only: instrument NOT included
Image coming soonN-003-033010-15-39USB-C / USB-A data cable 0.9 m (V2.0 instruments)
Spring-loaded bell probe holder (V-Block) – pencil probes, top cable exit models onlyF-000-700010-15-77Spring-loaded bell probe holder (V-Block) – pencil probes, top cable exit models only

For more on choosing the probe: probes for precision gauges and the guide How to choose the right ultrasonic transducer. For a quotation, please state the order code and quantity.

↑ Back to index

Documentation download

Official Dakota NDT technical documentation. The file opens in a new tab.

Download the PDFDatasheet Dakota PMX2-DL and PMX3-DL
Manufacturer’s official datasheet — PDF, in English

Standards and certifications

  • Factory calibration NIST-traceable and compliant with MIL-STD-45662A
  • Suitable for measurements to EN 14127 and ASTM E797

For the regulatory picture: NDT standards for thickness measurement.

Frequently asked questions

What is the difference between a precision gauge and a corrosion gauge?

The field of use and the type of probe. A corrosion gauge uses dual element probes and is optimised for thick, irregular walls, with resolution to the hundredth. A precision one such as the PMX2-DL uses single element, delay line and pencil probes, goes down to 0.001 mm of resolution and offers the Interface-Echo modes, which are essential on thin walls.

What is the minimum measurable thickness?

In Interface-Echo-Echo mode the range on steel starts from 0.152 mm; in Interface-Echo, measurement on plastics starts from 0.127 mm. The figures depend on the probe used and on the material.

What is the difference between the PMX2-DL and the PMX3-DL?

Only the display: greyscale on the PMX2-DL, colour AMOLED at 60 Hz on the PMX3-DL. Electronics, modes, range, gates and memory are identical. In greyscale the battery life is 35 hours against 12.

Can it also be used for corrosion measurements?

In contact Pulse-Echo mode the range reaches 9.14 metres on steel, so technically yes. For corrosion campaigns, however, the instruments of the CMX series are more suitable: they use dual element probes and offer the through-coating modes.

↑ Back to index

For a quotation or a trial on your own parts, contact our technical office.

Further information and contacts

Systems division and sensor integration partner : www.roder.it
More information about RODER SRL : about us
Contact for general information : info@roder.it
Contact for information about these products : info@roder.it


The information contained on this web page is provided for information purposes only. Although it has been compiled with the utmost care, it does not constitute a contractual offer or a binding commitment to supply. It may contain transcription, translation or typographical errors. For precise and up-to-date information, please contact our company directly.

Some images on this website have been intentionally generated using Artificial Intelligence (AI). This is because, for many applications and projects, it is not possible to disclose photographs of the installation or of the actual system due to confidentiality agreements, confidentiality clauses and non-disclosure agreements (NDAs).