NIST traceability and MIL-STD-45662A

Calibration blocks with certificate for ultrasonic thickness gauges

Traceability is a property of the measurement result, not a badge on the instrument. And the military standard that many specifications still cite was cancelled long ago.

  • VIM definition: documented, unbroken chain of calibrations
  • NIST does not certify the traceability of measurements it does not itself perform
  • MIL-STD-45662A, Rev. A of 1988: cancelled
  • Current references: ANSI/NCSL Z540-1 (then Z540.3), ISO 10012, ISO/IEC 17025

Calibration certificates often carry the phrase “NIST traceable”, and specifications just as often refer to MIL-STD-45662A. Both expressions deserve checking, because neither today means what those who write them usually intend.

Let us start from the definition. According to the VIM, the international vocabulary of metrology, metrological traceability is the “property of a measurement result whereby the result can be related to a reference through a documented unbroken chain of calibrations, each contributing to the measurement uncertainty”: the “property of a measurement result whereby the result can be related to a reference through a documented unbroken chain of calibrations, each contributing to the measurement uncertainty”.

NIST itself is explicit on two points. The first: “NIST does not define, specify, assure, or certify metrological traceability of the results of measurements except those that NIST itself provides”, that is, “NIST does not define, specify, assure, or certify metrological traceability of the results of measurements except those that NIST itself provides”. The second: “traceability is the property of the result of a measurement, not of an instrument or calibration report or laboratory”, that is, “traceability is the property of the result of a measurement, not of an instrument or calibration report or laboratory”.

MIL-STD-45662A: what happened

ItemItem
DocumentMILITARY STANDARD: CALIBRATION SYSTEM REQUIREMENTS, DoD
RevisionRev. A, 1 August 1988
StatusCancelled, marked as superseded by ISO 10012-1 and ANSI Z540-1
Date of supersessionIndication of 20 January 1995, reported by a single source
Current referencesANSI/NCSL Z540-1 (then Z540.3), ISO 10012, ISO/IEC 17025

What to write in a specification today

Calibration and verification procedure of an ultrasonic instrument

We think it right to say it plainly: MIL-STD-45662A is cancelled. Revision A is dated 1 August 1988 and is marked as superseded by ISO 10012-1 and ANSI Z540-1; the indication of supersession on 20 January 1995 comes from a single source and we report it as such, not as an established fact. Continuing to cite it in a purchase specification means requiring conformity to a document that is no longer in force.

The current references for calibration systems are ANSI/NCSL Z540-1 and the subsequent Z540.3, ISO 10012 and, for the competence of testing and calibration laboratories, ISO/IEC 17025. These are the ones to refer to when writing a new requirement.

As for the wording “NIST traceable”, the practical consequence of the two statements quoted above is simple: what makes a certificate defensible is not the name of an institute printed on the cover, but the documented chain of calibrations and the stated uncertainty. Our service is described on the Calibration and setting page.

Recommended instruments

ItemUse
Calibration blocksPhysical reference for verifying the measurement chain
Precision transducersProbes used in high-accuracy verifications
DakMasterData management and production of the measurement documentation
Dakota PMX2-DLUniversal precision gauge with data logger

Frequently asked questions

Is MIL-STD-45662A still in force?

No. Revision A, dated 1 August 1988, is cancelled and marked as superseded by ISO 10012-1 and ANSI Z540-1. There is also an indication of supersession on 20 January 1995, which we report as coming from a single source.

Which references should be cited instead?

The current references for calibration systems: ANSI/NCSL Z540-1 and the subsequent Z540.3, ISO 10012 and, for the competence of testing and calibration laboratories, ISO/IEC 17025.

What does “NIST traceable” really mean?

NIST states that it does not define, specify, assure or certify the metrological traceability of measurement results other than those it provides itself, and makes clear that traceability is a property of the result of a measurement, not of an instrument, a calibration report or a laboratory.

Back to Personnel and instrument qualification or write to us from the Contact page.

Further information and contacts

Systems division and sensor integration partner : www.roder.it
More information about RODER SRL : about us
Contact for general information : info@roder.it
Contact for information about these products : info@roder.it


The information contained on this web page is provided for information purposes only. Although it has been compiled with the utmost care, it does not constitute a contractual offer or a binding commitment to supply. It may contain transcription, translation or typographical errors. For precise and up-to-date information, please contact our company directly.

Some images on this website have been intentionally generated using Artificial Intelligence (AI). This is because, for many applications and projects, it is not possible to disclose photographs of the installation or of the actual system due to confidentiality agreements, confidentiality clauses and non-disclosure agreements (NDAs).